Login / Signup

Comparison of four different immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry assay for serum folate.

Lizi JinYouli LuXilian YiMeiwei ZhangJiangtao ZhangWeiyan ZhouJie ZengTianjiao ZhangChuanbao Zhang
Published in: Clinical chemistry and laboratory medicine (2022)
Assays for serum folate had good correlations with each other but lacked good agreement. The accuracy and consistency of assays for serum folate should be measured and assessed routinely. Standardization work to improve the accuracy of serum folate assays, such as the extension of traceability to reference methods or materials, calibration standardization efforts, and assay-adjusted cut-offs should be promoted.
Keyphrases
  • liquid chromatography tandem mass spectrometry
  • high throughput
  • simultaneous determination
  • ms ms
  • quality improvement