Login / Signup

Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract.

Shan MaRongyu GaoJing SunJun YangKai WenXiteng ChenFangyu ZhaoXinyan XuFang Tian
Published in: Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie (2023)
, it seems that the measurements should not be used interchangeably because of the wide LoA widths and the presence of significant differences among the devices. Similarly, due to significant differences and low ICCs, the PK, TK and TNP values obtained by IOLMaster 700, CASIA2, and Pentacam should not be used interchangeably.
Keyphrases
  • optical coherence tomography
  • diabetic retinopathy
  • high speed
  • mass spectrometry
  • cataract surgery
  • convolutional neural network
  • machine learning
  • high resolution
  • wound healing