Reconstructed spatial resolution and contrast recovery with Bayesian penalized likelihood reconstruction (Q.Clear) for FDG-PET compared to time-of-flight (TOF) with point spread function (PSF).

Julian Manuel Michael RogaschSaid SuleimanFrank HofheinzStephanie BluemelMathias LukasHolger AmthauerChristian Furth
Published in: EJNMMI physics (2020)
Q.Clear consistently improved reconstructed spatial resolution at high and medium SBR compared to PSF+TOF and OSEM+TOF, but only with beta = 150. However, this is at the cost of inferior SNR with Q.Clear150 compared to Q.Clear300/450 and PSF+TOF4/16/PSF+TOF2/17 while CRpeak for the small spheres did not improve considerably. This suggests that Q.Clear300/450 may be advantageous for the 3-ring detector configuration because the tradeoff between CR and SNR with Q.Clear300/450 was superior to PSF+TOF4/16, OSEM+TOF4/16, and OSEM+TOF2/17. However, it requires validation by systematic evaluation in patients at different activity and acquisition protocols.