Login / Signup

Two pumps or one pump? A comparison of human liver normothermic machine perfusion devices for transplantation.

Qiang LiuLuca Del PreteAhmed HassanDaniele PezzatiMary BilanciniGiuseppe D'AmicoTeresa Diago UsoKoji HashimotoFederico AucejoMasato FujikiKazunari SasakiChoon Hyuck David KwonBijan EghtesadCharles MillerCristiano Quintini
Published in: Artificial organs (2021)
Our data show that the 1-pump design provided the same hemodynamic output as the 2-pump design, with no additional hemolytic risk, but with the benefits of lower costs, easier transport and faster and simpler setting.
Keyphrases
  • deep learning
  • electronic health record
  • contrast enhanced
  • bone marrow
  • artificial intelligence
  • breast cancer risk