Utilization of CBCT to improve the delivery accuracy of Gamma Knife radiosurgery with G-frame.

Lindsey ClapsDamien MathewKathryn DusenberyMargaret ReynoldsYoichi Watanabe
Published in: Journal of applied clinical medical physics (2021)
The position differences reported by the registration module of LGP were within the accuracy limit of image registration for most clinical cases, but the errors could be larger in some cases. Therefore, we propose the following decision process. We do not advise position adjustment for G-frame based GKRS if the maximum displacement is less than 1 mm. When this limit is exceeded, however, another criterion should be applied to the decision making by considering the tumor size (or the treatment volume) together with the acceptable change of the tumor coverage.