Systematic Analysis of PTFE Monocusp Leaflet Design in a Patient-Based 3D in-Vitro Model of Tetralogy of Fallot.
Xiaoya ZhangShannen B KizilskiDominic P ReccoMartha D Chaillo LizarragaNicholas E KneierNoah E SchulzChristopher W BairdPeter E HammerDavid M HogansonPublished in: Cardiovascular engineering and technology (2023)
Balancing gradient, regurgitation, and oversizing for growth, a set of leaflet designs have been selected for pre-clinical evaluation. Designs with leaflet widths 140-160% in the 40% backwall model (110-120% in the 50% backwall), positioned at or 2 mm past the STJ, demonstrated the best results. The next stage of ex-vivo testing will additionally consider native RVOT distensibility, native leaflet interactions, and TAP characteristics.