Preparation of Dräger Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients.

Sebastian HeiderichChristian ThobenNils DennhardtTerence KraußRobert SümpelmannStefan ZimmermannMichael ReitzHenrik Rüffert
Published in: BMC anesthesiology (2021)
Both wash out protocols resulted in a significant reduction of trace gas concentrations. However, due to the complexity of the protocols and prolonged total preparation time, feasibility in clinical practice remains questionable. Especially when time is limited preparation of the anesthetic machines using ACF remain superior.