Login / Signup

Are tube fill volumes below 90% a rejection criterion for all coagulation tests?

Merve Sena OdabasiZeynep Mine Yalcinkaya Kara
Published in: Laboratory medicine (2024)
Sarstedt tubes meet quality standard specifications at a 70% fill rate for PT and aPTT and a 50% fill rate for fibrinogen. Comprehensive studies with larger populations are needed to accept these values as sample acceptance criteria for the laboratory.
Keyphrases
  • quality improvement