Head-to-head comparison of pressures during full cystometry, with clinical as well as in-depth signal-analysis, of air-filled catheters versus the ICS-standard water-filled catheters.
Peter F W M RosierPublished in: Neurourology and urodynamics (2021)
We confirm earlier studies that showed random differences at events between air-filled and water-filled pressures during clinical urodynamic testing and confirm that these are intrinsic but not systematic-and still incompletely explained-offset-baseline differences. We determined on closer full measurement analysis after equalizing, that both systems are similar in displaying urodynamic pressure variations and amplitudes. We also confirm that both systems require awareness of intrinsic measurement properties during urodynamic testing and especially may necessitate adjustment of pressure offsets into a quantitative diagnosis of a urodynamic test.