Image quality assessment of three cone beam computed tomography scanners-an analysis of the visibility of anatomical landmarks.

Elisa BaumannMichael M BornsteinMichel DalstraCarlalberta VernaDorothea C Dagassan-Berndt
Published in: European journal of orthodontics (2022)
CBCT scans performed for orthodontic purposes using a large FoV with reduced parameters (400 µm, 2 to 4 mA and low dose protocols) are acceptable for visualization of large anatomical structures. Further lowering these parameters will not be sufficient to view small anatomical structures. Orthodontic indications will have to define specific anatomical structures to choose adequate scanning protocols to reduce dose and ensure reliable diagnostic visibility.