Login / Signup

Transcatheter aortic valve deployment influences neo-sinus thrombosis risk: An in vitro flow study.

Immanuel David Madukauwa-DavidVahid SadriNorihiko KamiokaPrem A MidhaVrishank RaghavJohn N OshinskiRahul SharmaVasilis BabaliarosShelly Singh-Gryzbon
Published in: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2019)
Neo-sinus flow stasis is influenced by THV expansion and implant depth (CoreValve). The 10% valve under-deployment (oversizing) may facilitate reduced flow stasis in the neo-sinus with minimal increase in pressure gradients. This strategy may be helpful for patient anatomies, which are in-between transcatheter valve sizes.