Login / Signup

Diagnostic Performance of the PalmScan VF2000 Virtual Reality Visual Field Analyzer for Identification and Classification of Glaucoma.

Vijay ShettyPrachi SankheSuhas S HaldipurkarTanvi HaldipurkarRita DhamankarPriyanka KashelkarDhruven ShahParesh MhatreManinder Singh Setia
Published in: Journal of ophthalmic & vision research (2022)
The instrument is 100% sensitive and specific in detection of glaucoma. However, among patients with glaucoma, there is a relatively high proportion of misclassification of severity of glaucoma. Thus, although useful for screening of glaucoma, it cannot replace the HumphreyⓇ Field Analyzer for the clinical management in its current form.
Keyphrases
  • optic nerve
  • virtual reality
  • cataract surgery
  • machine learning
  • deep learning