Evaluating on-board kVCT- and MVCT-based dose calculation accuracy using a thorax phantom for helical tomotherapy treatments.
Riley C TegtmeierWilliam S FerrisRuiming ChenJessica MillerJohn BayouthWesley S CulbersonPublished in: Biomedical physics & engineering express (2023)
Accurate dose calculations (median errors of < +/-1%) using a thorax phantom simulating realistic patient geometry and scatter conditions can be achieved with images acquired with a helical kVCT system on a helical tomotherapy unit. This accuracy is considerably improved relative to that achieved with the MV-based approach. In a clinical setting, careful consideration should be made when selecting appropriate kVCT imaging parameters for this process as dose calculation accuracy was observed to vary with both parameter selection and treatment.