Login / Signup

Evaluation of iliac screw, S2 alar-iliac screw and laterally placed triangular titanium implants for sacropelvic fixation in combination with posterior lumbar instrumentation: a finite element study.

Gloria CasaroliLuca Maria SconfienzaRuchi ChandeDerek LindseyAli MesiwalaScott YerbyMarco Brayda-Bruno
Published in: European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society (2019)
Implant stresses after S2AI and SI fixations were lower than those attributable to IL. Therefore, pedicle screws and rods may have a lower risk of mechanical failure when coupled with sacropelvic fixation via S2AI or triangular titanium implants, although the risk of clinical loosening remains an area of further investigation. These slides can be retrieved under Electronic Supplementary Material.