Quantitative validation of MRI mapping of cerebral venous oxygenation with direct blood sampling: A graded-O2 study in piglets.

Dengrong JiangRaymond C KoehlerXiuyun LiuEwa KulikowiczJennifer K LeeHanzhang LuPeiying Liu
Published in: Magnetic resonance in medicine (2021)
The values of Yv measured by TRUPC and aTRUPC were in good agreement with blood gas oximetry. These findings suggest that TRUPC and aTRUPC can provide accurate quantifications of Yv in major cerebral veins.
Keyphrases
  • high resolution
  • subarachnoid hemorrhage
  • magnetic resonance imaging
  • contrast enhanced
  • cerebral ischemia
  • inferior vena cava
  • pulmonary embolism
  • high density
  • carbon dioxide