Design and validation of an automated dual-arm instrumented intravaginal dynamometer.
Catriona S CzyrnyjMarie-Eve BerubeEric LanteigneAna BrennanYousef BaderDmitry LomovtsevMegan VandermolenSamantha BoucherLayla MitriLinda McLeanPublished in: Neurourology and urodynamics (2021)
This IVD design demonstrates valid load measurement and speed control during bench testing. Active and passive forces measured are consistent with the literature. With dual instrumented arms, this device allows for further investigation into the source of measured vaginal closure forces.