Reproducibility, Repeatability, and Correlation of Central Corneal Thickness Measurement with the Pentacam Scheimpflug System and Ultrasound Pachymetry.

Yolanda Macias-RodriguezEugenia M Ramos-DávilaRaul Eduardo Ruiz-LozanoJose V Reyes-ArenaI Jocelyn Rivera-AlvaradoJulio C Hernandez-CamarenaAlejandro Rodriguez-Garcia
Published in: Klinische Monatsblatter fur Augenheilkunde (2022)
The Pentacam HR device provides reliable operator-independent estimates of CCT. Right eyes exhibited a tendency to overestimate with the Pentacam HR. We suspect this difference is due to USP underestimation related to patients' position while performing the study. In clinically relevant scenarios, performing a third measurement and cautiously measuring right eyes can provide higher accuracy.