Login / Signup

A high-throughput LC-MS/MS method for determination of flunarizine in human plasma: Pharmacokinetic study with different doses.

Abhaysingh BhadoriyaPriyanka A ShahPranav S ShrivastavMallika SanyalManish S Yadav
Published in: Biomedical chromatography : BMC (2019)
A high-throughput and sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method has been developed and validated for the determination of flunarizine in human plasma. Liquid-liquid extraction under acidic conditions was used to extract flunarizine and flunarizine-d8 from 100 μL human plasma. The mean extraction recovery obtained for flunarizine was 98.85% without compromising the sensitivity of the method. The chromatographic separation was performed on Hypersil Gold C18 (50 × 2.1 mm, 3 μm) column using methanol-10 mm ammonium formate, pH 3.0 (90:10, v/v) as the mobile phase. A tandem mass spectrometer (API-5500) equipped with an electrospray ionization source in the positive ion mode was used for detection of flunarizine. Multiple reaction monitoring was selected for quantitation using the transitions, m/z 405.2 → 203.2 for flunarizine and m/z 413.1 → 203.2 for flunarizine-d8. The validated concentration range was established from 0.10 to 100 ng/mL. The accuracy (96.1-103.1%), intra-batch and inter-batch precision (CV ≤ 5.2%) were satisfactory and the drug was stable in human plasma under all tested conditions. The method was used to evaluate the pharmacokinetics of 5 and 10 mg flunarizine tablet formulation in 24 healthy subjects. The pharmacokinetic parameters Cmax and AUC were dose-proportional.
Keyphrases