Login / Signup

Analysis of the Circulating Tumor Cell Capture Ability of a Slit Filter-Based Method in Comparison to a Selection-Free Method in Multiple Cancer Types.

Hidenori TakagiLiang DongMorgan D KuczlerKara A LombardoMitsuharu HiraiSarah R AmendKenneth J Pienta
Published in: International journal of molecular sciences (2020)
Circulating tumor cells (CTCs) are a promising biomarker for cancer liquid biopsy. To evaluate the CTC capture bias and detection capability of the slit filter-based CTC isolation platform (CTC-FIND), we prospectively compared it head to head to a selection-free platform (AccuCyte®-CyteFinder® system). We used the two methods to determine the CTC counts, CTC positive rates, CTC size distributions, and CTC phenotypes in 36 patients with metastatic cancer. Between the two methods, the median CTC counts were not significantly different and the total counts were correlated (r = 0.63, p < 0.0001). The CTC positive rate by CTC-FIND was significantly higher than that by AccuCyte®-CyteFinder® system (91.7% vs. 66.7%, p < 0.05). The median diameter of CTCs collected by CTC-FIND was significantly larger (13.0 μm, range 5.2-52.0 vs. 10.4 μm, range 5.2-44.2, p < 0.0001). The distributions of CTC phenotypes (CK+EpCAM+, CK+EpCAM- or CK-EpCAM+) detected by both methods were similar. These results suggested that CTC-FIND can detect more CTC-positive cases but with a bias toward large size of CTCs.
Keyphrases
  • circulating tumor cells
  • circulating tumor
  • stem cells
  • high throughput
  • protein kinase
  • squamous cell
  • peripheral blood
  • cell free
  • young adults
  • loop mediated isothermal amplification