Login / Signup

Analytical performance of the digital morphology analyzer Sysmex DI-60 for body fluid cell differential counts.

Sumi YoonHye Ryoun Kim
Published in: PloS one (2023)
This is the first study to evaluate the performance of DI-60 compared with XN-350 and manual counting for BF cell differential counts. DI-60 showed reliable performance with CSF, pleural fluid, and ascites samples. For BF cell differential counts, DI-60 may be a better option than XN-350 and could be used for screening purposes in understaffed laboratories. To improve the hematology workflow for BF cell differential counting, the DM analyzer needs to be optimized by taking into account the laboratory situation and unmet needs, and the clinical laboratory needs to establish criteria for verification and manual slide review.
Keyphrases
  • single cell
  • cell therapy
  • biofilm formation
  • stem cells
  • type diabetes
  • metabolic syndrome
  • escherichia coli
  • staphylococcus aureus
  • skeletal muscle
  • pseudomonas aeruginosa
  • electronic health record