Assessment of Heating on Titanium Alloy Cerebral Aneurysm Clips during 7T MRI.

S TsutsuiTsuyoshi MatsudaK TakedaMakoto SasakiYoshitaka KuboK SettaShunrou FujiwaraKohei ChidaKuniaki Ogasawara
Published in: AJNR. American journal of neuroradiology (2022)
Temperature increases on the right-angled and short straight clips remained below the regulatory limit during 7T MR imaging, but temperature increases on the 45° angled, straight fenestrated, and long straight clips exceeded this limit.