Impact of Patient Interface Diameter and Vacuum Level on Suction Stability Using a Flat Applanating Interface for Femtosecond Laser-Assisted LASIK.

Siegfried MariacherMartina MariacherAnna Maria SeuthePeter SzurmanBerthold SeitzKarl Thomas Boden
Published in: Current eye research (2019)
The vacuum level and the diameter of the PI showed significant impact on suction stability measured as break-away force during flat applanating docking for corneal and refractive surgery. Break-away forces were inversely related to the PI diameter and directly to the vacuum level. Increasing the PI diameter by one step or reducing vacuum by 100 mbar resulted in a comparable decrease of break-away forces. Therefore, the surgeon could potentially maintain stabile suction by increasing vacuum when choosing a larger PI diameter to obtain a larger treatment zone. Furthermore, reduction of the PI diameter and/or increasing vacuum could help improving suction stability in situations of poor suction and in situations with increased risk of suction loss during LASIK.