Login / Signup

Experimental characterisation of the magnetic field correction factor,kB⃗,for Roos chambers in a parallel MRI-linac.

Jarrad BeggUrszula JelenPaul J KeallGary LineyLois C Holloway
Published in: Physics in medicine and biology (2022)
Objective. Reference dosimetry on an MRI-linac requires a chamber specific magnetic field correction factor,kB⃗.This work aims to measure the correction factor for a parallel plate chamber on a parallel MRI-linac. Approach. kB⃗is defined as the ratio of the absorbed dose to water calibration coefficient in the presence of the magnetic field,ND,wB⃗relative to that under 0 T conditions,ND,w0T.kB⃗was measured via aND,wtransfer to a field chamber at each magnetic field strength from a chamber with knownND,wandkB⃗.This was achieved on the parallel MRI-linac by moving the measurement set-up between a high magnetic field strength region at the MRI-isocentre and a low magnetic field strength region at the end of the bore whilst maintaining consistent set-up and scatter conditions. Three PTW 34001 Roos chambers were investigated as well as a PTW 30013 Farmer used to validate methodology. Main Results. The beam quality used for the measurements ofkB⃗was TPR 20 / 10  = 0.632. ThekB⃗for the PTW Farmer chamber at 1 T on a parallel MRI-linac was 0.993 ± 0.013 ( k  = 1). The averagekB⃗factor measured for the three Roos chambers on a 1 T parallel MRI-linac was 0.999 ± 0.014 ( k  = 1). Significance. The results presented are the first measurements ofkB⃗for a Roos chamber on a parallel MRI-linac. The Roos chamber results demonstrate the potential for the chamber as a reference dosimeter in parallel MRI-linacs.
Keyphrases
  • contrast enhanced
  • magnetic resonance imaging
  • diffusion weighted imaging
  • computed tomography