Login / Signup

Comprehensive SPECT/CT system characterization and calibration for 177Lu quantitative SPECT (QSPECT) with dead-time correction.

Andrea FrezzaCorentin DesportCarlos UribeWei ZhaoAnna CellerPhilippe DesprésJean-Mathieu Beauregard
Published in: EJNMMI physics (2020)
Serial planar acquisitions of sources in air using an activity range covering the full operational capacity of the SPECT/CT system, with multiple energy windows for wide-spectrum count rate monitoring, and followed by simultaneous determination of CF and τ using a single equation derived from the paralysable model, constitutes a practical method to enable accurate dead-time-corrected QSPECT imaging in a post-177Lu radionuclide therapy setting.
Keyphrases