Spectral correction of light emitting diodes enables accurate hydration ratio calculation using narrowband diffuse reflectance spectroscopy.

Jesse H LamJeonghun KimKelsey J TuDongbin KimSehwan Kim
Published in: Journal of biomedical optics (2023)
errors increased with the spectral bandwidth of LEDs but could be corrected using SC. Likewise, in emulsions, we found an average error of 8.7% (maximum error 14%) if SC was not used. By contrast, applying SC reduced the average error to 2.2% (maximum error of 6.4%). We show that despite utilizing multiple, spectrally broad, and overlapping LEDs, SC was still able to restore the performance of our narrowband method, making it comparable to a much larger full broadband system.