Performance evaluation of a novel platelet count parameter, hybrid platelet count, on the BC-780 automated hematology analyzer.

Eakachai PrompetcharaChalisa ParnsamutAngkana ChirapanurukChutitorn Ketloy
Published in: Clinical chemistry and laboratory medicine (2023)
The performance of PLT-H in the BC-780 met the manufacturer's specifications. PLT-H exhibits better reproducibility than did PLT-I, correlates well with the PLT-O for thrombocytopenic samples and demonstrates good anti-interference ability. PLT-H counting is therefore recommended as a zero-cost alternative platelet counting method for platelet interference samples in clinical settings.