Login / Signup

Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies.

Joseph J RuzbarskyIshaan SwarupMatthew R GarnerKathleen N MeyersFolorunsho Edobor-OsulaRoger F WidmannDavid M Scher
Published in: HSS journal : the musculoskeletal journal of Hospital for Special Surgery (2019)
Plate B failed at a higher axial load in biomechanical testing, likely related to differences in its design. The addition of a proximal screw did not increase the axial loading properties of the blade plate construct but did increase the pullout strength by a factor of 12. These results may be used to influence implant selection and post-operative rehabilitation following proximal femoral osteotomies in children.
Keyphrases
  • finite element analysis
  • young adults
  • finite element