Physical activity detection in patients with intracardiac leadless pacemaker.

Zsolt BariMáté VámosPeter BogyiDwight ReynoldsTodd SheldonDedra H FaganGabor Zoltan Duray
Published in: Journal of cardiovascular electrophysiology (2018)
The three-axis accelerometer-based rate adaptive pacing feature proved to be feasible after manual selection of an adequate activity vector. Vector testing in Micra patients with chronotropic incompetence appears to be beneficial compared with the use of nominal Vector 1.