Login / Signup

Validation of a Molecular Diagnostic Test for Circulating Tumor DNA by Next-Gen Sequencing.

Sandra V FernandezYin Fei TanShilpa RaoPatricia FittipaldiFathima SheriffHossein BorghaeiEfrat DotanJennifer S WinnMartin J EdelmanJoseph TreatJulia JuddR Katherine AlpaughY Lynn WangJian Q YuMariusz WasikDon A Baldwin
Published in: International journal of molecular sciences (2023)
A modified version of the PGDx elio TM Plasma Resolve assay was validated as a laboratory-developed test (LDT) for clinical use in the Molecular Diagnostics Laboratory at Fox Chase Cancer Center. The test detects single nucleotide variants (SNVs) and small insertions and deletions (indels) in 33 target genes using fragmented genomic DNA extracted from plasma. The analytical performance of this assay was assessed with reference standard DNA and 29 samples from cancer patients and detected 66 SNVs and 23 indels. Using 50 ng of input DNA, the sensitivity was 95.5% to detect SNVs at 0.5% allele frequency, and the specificity was 92.3%. The sensitivity to detect indels at 1% allele frequency was 70.4%. A cutoff of 0.25% variant allele frequency (VAF) was set up for diagnostic reporting. An inter-laboratory study of concordance with an orthologous test resulted in a positive percent agreement (PPA) of 91.7%.
Keyphrases
  • circulating tumor
  • cell free
  • circulating tumor cells
  • single molecule
  • high throughput
  • copy number
  • genome wide
  • emergency department
  • mass spectrometry
  • dna methylation
  • transcription factor