Label-Free, Direct Measurement of Protein Concentrations in Turbid Solutions with a UV-Visible Integrating Cavity Absorbance Spectrometer.
Nicholas R LarsonYangjie WeiC Russell MiddaughPublished in: Analytical chemistry (2018)
Protein-particle conjugates and mixtures have been investigated extensively for their diverse applications in biotechnology. However, general methods to measure protein concentration of protein-particle solutions are lacking. Typically, proteins in turbid solutions require separation or staining with another chromophore to quantitate their concentration. Here we demonstrate a label-free, direct approach to measure protein concentrations in turbid solutions using a UV-vis integrating cavity absorbance spectrometer. Three systems are used to test the ability to measure accurate protein concentrations: proteins adsorbed to Alhydrogel, proteins in solution with gold nanoparticles, and proteins encapsulated within polymeric microspheres. Protein concentrations in each of the three protein-particle systems were successfully quantified using a calibration curve created from the absorbance at 280 nm.