Measurement of corneal thickness using Pentacam HR versus Nidek CEM-530 specular microscopy.
Maddalena De BernardoPalmiro CornettaGiuseppe MarottaGiulio SalernoIlaria De PascaleNicola RosaPublished in: The Journal of international medical research (2019)
These findings indicate that the CEM-530 produces a thicker corneal measurement than the Pentacam HR. The herein-proposed correcting factors are needed to reliably compare these devices.