Failure of a Rotation Control Gamma 3 Lag Screw Used to Treat a Trochanteric Fracture.
Kyungho ChoiYongtae KimShicheng ZhouJi Hyo HwangPublished in: Hip & pelvis (2018)
Gamma 3 rotation control lag screws (U-blade) are particularly useful when treating rotational and unstable fractures of the proximal femur. A 93-year-old woman who underwent closed reduction of a trochanteric fracture and internal fixation with a Gamma 3 nail rotation control lag screw. The patient presented with metal failure and U-blade bending following a fall occurring 4 weeks after surgery. Here, we present a case report summarizing removal of the failed lag screw.