A cluster of three extrapulmonary Mycobacterium abscessus infections linked to well-maintained water-based heater-cooler devices.
Jessica L SeidelmanArthur W BakerSarah S LewisBobby Glenn WarrenAaron BarrettAmanda GravesCarly KingBonnie TaylorJill EngelDesiree BonnadonnaCarmelo MilanoRichard J WallaceMatthew StiegelDeverick J AndersonBecky A SmithPublished in: Infection control and hospital epidemiology (2023)
to 3 patients. Design modifications to better contain aerosols or filter exhaust during device operation are needed to prevent NTM transmission events from water-based HCDs.