Effects of Angular Resolution and b Value on Diffusion Tensor Imaging in Knee Joint.

Qi ZhaoRees P RidoutJikai ShenNian Wang
Published in: Cartilage (2021)
To obtain consistent DTI outputs and tractography in knee joint, the scan may require a proper b value (ranging from 500 to 1500 s/mm2) and sufficient angular resolution (>14) with signal-to-noise ratio >10.