Evaluation of the differences in pressure applied to the vessel wall by different types of balloon remodeling microcatheters in an experimental model.

Mizuka IkezawaTakashi IzumiMasahiro NishihoriTetsuya TsukadaYosuke TamariYoshio ArakiKinya YokoyamaKenji UdaShunsaku GotoAsuka E KroppTakafumi OtsukaNaoki KatoMizuki NakanoRyuta Saito
Published in: Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences (2021)
Pressure differed across balloon remodeling microcatheters. In addition, vessel shape and diameter, and hole size, affected the results. Our findings can help select balloon remodeling microcatheters.