Comparing the accuracy of the da Vinci Xi and da Vinci Si for image guidance and automation.
James M FergusonE Bryn PittAlan KuntzJosephine GrannaNicholas L KavoussiNaren NimmagaddaEric J BarthStanley Duke HerrellRobert J WebsterPublished in: The international journal of medical robotics + computer assisted surgery : MRCAS (2020)
The Xi can achieve submillimetric average error. Calibration improves accuracy, but excellent baseline accuracy of the Xi means that calibration may not be needed for some applications. Importantly, the external tracking systems needed to account for setup joint error in the Si are no longer required with the Xi.