Login / Signup

Evaluation of a Hydrogel-Based Diagnostic Approach for the Point-of-Care Based Detection of Neisseria gonorrhoeae.

Sumudu R PereraAli TaheriNurul H KhanRajinder P PartiStephanie StefuraPauline SkibaJason P AckerIrene MartinAnthony J KusalikJo-Anne R Dillon
Published in: Antibiotics (Basel, Switzerland) (2018)
Eleven primer pairs were developed for the identification of Neisseria gonorrhoeae. The sensitivity and specificity of these primers were evaluated by Real Time (RT)-PCR melt curve analyses with DNA from 145 N. gonorrhoeae isolates and 40 other Neisseria or non-Neisseria species. Three primer pairs were further evaluated in a hydrogel-based RT-PCR detection platform, using DNA extracted from 50 N. gonorrhoeae cultures. We observed 100% sensitivity and specificity in the hydrogel assay, confirming its potential as a point-of-care test (POCT) for N. gonorrhoeae diagnosis.