Login / Signup

Subcutaneous ICD screening with the Boston Scientific ZOOM programmer versus a 12-lead ECG machine.

Shu C ChangKristen K PattonMelissa R RobinsonJeanne E PooleJordan M Prutkin
Published in: Pacing and clinical electrophysiology : PACE (2018)
There can be occasional disagreement in S-ICD patient screening between an ICD programmer and ECG machine, all of whom passed with the ECG machine but failed using the programmer. On a per lead basis, the ECG machine passes more subjects. It is unknown what the inappropriate shock rate would be if an S-ICD was implanted. Clinical judgment should be used in borderline cases.
Keyphrases
  • heart rate variability
  • deep learning
  • heart rate
  • case report
  • machine learning