Development and validation of HPLC-MS/MS method for the determination of lixivaptan in mouse plasma and its application in a pharmacokinetic study.

Adnan A KadiHaitham AlrabiahMohamed W AttwaSabry AttiaGamal A E Mostafa
Published in: Biomedical chromatography : BMC (2017)
The aim of the present study was to develop a simple, sensitive and accurate liquid chromatography-electrospray ionization tandem mass spectrometry (ESI-MS/MS) method for the determination of lixivaptan (LIX) in mouse plasma using vildagliptin as the internal standard (IS). A precipitation procedure was used for the extraction of LIX and vildagliptin from mouse plasma. Chromatographic separation of LIX was achieved using a C18 analytical column (50 × 2.1 mm, 1.8 μm) at 25°C. The mobile phase comprised acetonitrile and ammonium formate (10 mm, pH 3.1; 40:60, v/v) pumped at a flow rate of 0.3 mL min-1 . A tandem mass spectrometer with an electrospray ionization source was used to perform the assay. Quantification of LIX at m/z 290 → 137 and IS at 154 → 97 was attained through multiple reaction monitoring. The investigated method was authenticated following the bio-analytical method of validation guidelines of the US Food and Drug Administration. The developed method showed a good linearity over the concentration range from 5 to 500 ng mL-1 , and the calibration curve was linear (r = 0.9998). The mean recovery of LIX from mouse plasma was 99.2 ± 0.68%. All validation parameters for LIX were within the levels required for acceptance. The proposed method was effectively used for a pharmacokinetic study of LIX in mouse plasma.