An automated and portable antimicrobial susceptibility testing system for urinary tract infections.

Kuo-Wei HsuWen-Bin LeeHuey-Ling YouMel S LeeGwo-Bin Lee
Published in: Lab on a chip (2021)
Urinary tract infections (UTIs) are bacterial infections that 1) commonly affect females, 2) can pose high risks to impair kidney function, 3) are often treated with broad-spectrum antibiotics, and 4) are associated with high recurrence rates due to the evolution of drug-resistant strains. To choose the appropriate antibiotic, the minimum inhibitory concentration (MIC) among a panel of antibiotics should be determined before administration to avoid inadequate dosing or use of wrong antibiotics. To meet with the unmet needs, we developed a bead-based method for bacterial preconcentration with capture rates ranging from 20-50% and then automatically performed on-chip AST on an automated device which was composed of a pneumatic control module, a temperature control module and a chip image processing module. The developed portable system was capable of automatically conducting AST and MIC measurements using urine samples (via image analysis) in only 4.5-9 h and tested on four common UTIs bacterial strains. This compact system may therefore be promising for point-of-care personalized medicine in the near future.