Comparing ocular biometry and autorefraction measurements from the Myopia Master with the IOLMaster 700 and the Huvitz HRK-8000A autorefractor.

Hilde Røgeberg PedersenEllen SvarverudLene A HagenStuart J GilsonRigmor C Baraas
Published in: Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) (2023)
The LoA for measurements of SER, CR and AL when comparing the Myopia Master with the HRK-8000A and the IOLMaster 700 were wider than deemed acceptable for making direct comparisons. This indicates that the instruments cannot be used interchangeably in clinical practice or research.