Hybrid Approach for Embolized Right Ventricular Outflow Tract Stent.
Sudesh PrabhuRishi TiwariSiddhant MehraRiyan ShettyPrakash RamachandraShreesha MaiyaPublished in: Innovations (Philadelphia, Pa.) (2020)
Displacement and embolization of the stent is one of the major complications of right ventricular outflow tract (RVOT) stenting. Since embolized stents cannot be retrieved percutaneously as they cannot be crimped into pre-stenting state (unlike duct occluders and septal devices, which can be pulled back into the sheath), surgery remains the gold standard treatment. We describe a hybrid approach for the retrieval of the embolized RVOT stent, which will decrease morbidity when compared to the standard surgical approach that is otherwise required.