Evaluation and reduction of magnetic resonance imaging artefacts induced by distinct plates for osseous fixation: an in vitro study @ 3 T.
Carsten RendenbachMax SchoellchenJulie BueschelTobias GauerJan SedlacikDaniel KutznerPekka K VallittuMax HeilandRalf SmeetsJens FiehlerSusanne SiemonsenPublished in: Dento maxillo facial radiology (2018)
Polylactide, GFRC and magnesium plates produce less susceptibility artefacts in MRI compared to TI, while the dimensions of TI plates directly influence artefact extension. Slice encoding for metal artefact reduction and ultrashort echo time significantly reduce metal artefacts at the expense of scan time or image resolution.