Framework for evaluating photon-counting detectors under pile-up conditions.

David LeiboldStefan J van der SarMarlies C GoordenDennis R Schaart
Published in: Journal of medical imaging (Bellingham, Wash.) (2024)
The proposed framework permits quantifying the spectral and spatial performance of PCDs under clinically realistic conditions at a given operating point. The presented example illustrates why PCDs should not be analyzed assuming that they are linear systems. The framework can, for example, be used to guide the development of PCDs and PCD-based systems. Furthermore, it can be applied to adapt commonly used measures, such as the modulation transfer function, to non-linear PCDs.