Login / Signup

Rapid Mass Spectrometric Calibration and Standard Addition Using Hydrophobic/Hydrophilic Patterned Surfaces and Discontinuous Dewetting.

Matthias HermannPrashant AgrawalChang LiuJ C Yves Le BlancThomas R CoveyRichard David Oleschuk
Published in: Journal of the American Society for Mass Spectrometry (2022)
The rapid calibration chip (RCC) is a device that uses the fast and reproducible wetting behavior of hydrophilic/hydrophobic patterned surfaces to confine a series of differently sized droplets on a substrate to obtain a calibration curve. Multiple series of droplets can be formed within seconds by dipping an RCC into a calibration solution. No pipetting, sequential droplet deposition, or advanced equipment is required. The performance and reproducibility of RCCs were evaluated with an electrospray ionization triple-quadrupole mass spectrometer equipped with a liquid microjunction-surface sampling probe (LMJ-SSP) that allows for fast sampling of surfaces. Using circular hydrophilic areas with diameters ranging from 0.25 to 2.00 mm, liquid volumes of 4.6-70.6 nL could be deposited. Furthermore, the use of a second hydrophobic/hydrophilic patterned transfer chip can be used to add internal standard solutions to each calibration spot of the RCC, allowing to transfer a liquid volume of 22.5 nL.
Keyphrases