T-Wave Oversensing with Contemporary Implantable Cardioverter-Defibrillators.
Marc StrikSylvain PlouxRomain EschalierPierre MondolyLeslie FontagneF Daniel RamirezMichel HaïssaguerrePierre BordacharPublished in: Journal of cardiovascular development and disease (2023)
On review of 24,379 transmitted episodes from 2407 patients with endovascular ICDs, we found no case of TWOS with Boston Scientific devices, whereas TWOS was not uncommon with Medtronic devices. However, the risk of inappropriate therapy with Medtronic ICDs was very low (0.1%) due to the often intermittent nature of this phenomenon, the morphology discriminator, and the anti-TWOS algorithm.