Performance of commercially-available cholesterol self-tests.
Steef KurstjensEugenie GemenSelina WalkTjin NjoJohannes KrabbeKarlijn GijzenMarc Glm ElisenRon KustersPublished in: Annals of clinical biochemistry (2021)
The average plasma cholesterol concentration was 5.2 ± 1.2 mmol/L. The mean absolute relative difference (MARD) of the five cholesterol self-tests ranged from 6 ± 5% (Accutrend Plus) to 20 ± 12% (Mylan Mytest). The Accutrend Plus cholesterol meter showed the best diagnostic performance with a 92% sensitivity and 89% specificity. The Qucare and Mission 3-in-1 are able to measure HDL concentrations and can thus provide a cholesterol:HDL ratio. The Passing-Bablok regression analyses for the ratio showed poor performance in both self-tests (Mission 3-in-1: y = 1.62x-1.20; Qucare: y = 0.61x + 1.75). The Accutrend Plus is unable to measure the plasma high-density lipoprotein concentration.Conclusions/interpretation: The Accutrend Plus cholesterol meter (Roche) had excellent diagnostic and analytic performance. However, several of the commercially-available self-tests had considerably poor accuracy and diagnostic performance and therefore do not meet the required qualifications, potentially leading to erroneous results. Better regulation, standardization and harmonization of cholesterol self-tests is warranted.