Comparison of the performance of battery-operated fluid warmers.
Amit LehaviAvraham YitzhakRefael JarassyRami HeizlerYeshayahu Shai KatzAeyal RazPublished in: Emergency medicine journal : EMJ (2018)
We found significant differences between the fluid warmers: the use of the Buddy Lite should be limited to moderate input temperature and low flow rates. The use of the Thermal Angel is limited to low volumes due to battery capacity and low output temperature at extreme conditions. The Warrior provides the best warming performance at high infusion rates, as well as low input temperatures, and was able to warm the largest volumes in these conditions.