Specific absorption rate and temperature in neonate models resulting from exposure to a 7T head coil.
Shaihan J MalikJeffrey W HandRyan SatnarineAnthony N PriceJoseph V HajnalPublished in: Magnetic resonance in medicine (2021)
This study of two neonate models cannot quantify the variability expected within a larger population. Likewise, the use of age-adjusted dielectric properties have a significant effect, but while their use is well motivated by literature, there is uncertainty in the true dielectric properties of neonatal tissue. Nevertheless, the main finding is that unlike at lower field strengths, operational limits for 7T neonatal MRI using an adult head coil should be more conservative than limits for use on adults.