Comparison of Six Automated Treponema-Specific Antibody Assays.
Borae G ParkJihoon G YoonJohn Hoon RimAnna LeeHyon-Suk KimPublished in: Journal of clinical microbiology (2015)
Six different Treponema (TP)-specific immunoassays were compared to the fluorescent treponemal antibody absorption (FTA-ABS) test. A total of 615 samples were tested. The overall percent agreement, analytical sensitivity, and analytical specificity of each assay compared to the FTA-ABS test were as follows: Architect Syphilis TP, 99.2%, 96.8%, and 100%; Cobas Syphilis, 99.8%, 99.4%, and 100%; ADVIA Centaur Syphilis, 99.8%, 99.4%, and 100%; HISCL Anti-TP assay kit, 99.7%, 98.7%, and 100%; Immunoticles Auto3 TP, 99.0%, 97.5%, and 99.6%; Mediace TPLA, 98.0%, 98.1%, and 98.0%. All results that were discrepant between the TP-specific assays were associated with samples from noninfectious cases (11 immunoassay false positives and 7 from previous syphilis cases). Our study demonstrated that TP-specific immunoassays generally showed high sensitivities, specificities, and percentages of agreement compared to FTA-ABS, with rare cases of false-positive or false-negative results. Therefore, most TP-specific immunoassays are acceptable for use in screening for syphilis. However, it is important to perform a thorough review of a patient's clinical and treatment history for interpreting the results of syphilis serology.